Peltron - przetworniki ci�nienia, czujniki lvdt

Home About us Pressure Displacement Universal devices Why it is good.. Contact

 

Differential pressure transmitters


Larger
NPXD series
Product description:
NPXD series is a group of transducers from NPX series equipped in two pressure connectors, used to measure differential pressure. Other parameters like in NPX series

Technical parameters:
Measure
ranges [bar]
▪ 0÷0,050 minimal
▪ 0÷20 maximal
Power supply▪ 10÷36  VDC for o. s. 4÷20 mA
▪ 15÷30  VDC for o. s. 0÷10V, 0÷5 V, 1÷5 V
▪ 12÷30  VDC for o. s. 0÷5 V (option)
▪ 4,5÷12 VDC for o. s. 0÷3 V (low-power)
Output
signals
▪ 4÷20 mA Rload ≤(Ub-10)V/0,02 A
▪ 0÷10 V    Rload ≤10 kΩ
▪ 0÷5 V      Rload ≤10 kΩ
▪ 0÷3 V      Rload ≤10 kΩ

* There is possibility to make transducers with any volatage output signal lesser than 0÷10V (fe. 0÷4V, 2÷8V and so on.)
Allowable overloading
4 x range
Basic error [%MR]▪ ≤ 0,5
▪ ≤ 0,25
Pressure connectors▪ M20x1,5 / M12x1,5 outer thread
▪ G¼" / G½" outer thread
▪ St. B¼" / St. B½" inner thread
▪ others to agree
Ingress ProtectionIP 65
Applicationheat networks, water systems

                               Look at technical specification

Larger
PXD series
Product description:
PXD series - is a group of transducers from PX series equipped in two pressure connections, used in measuring of differential pressure. Other parameters like in PX series.

Technical parameters:
Measure
ranges [bar]
▪ 0÷0,050 minimal
▪ 0÷20 maximal
Power supply15÷36V DC
Output signals▪ 4÷20 mA
▪ 0÷20 mA
▪ 20÷0 mA
▪ 0÷10 V
Basic error [%MR]▪ ≤ 0,5
▪ ≤ 0,25
Pressure connectors▪ M20x1,5 / M12x1,5 outer thread
▪ G¼" / G½" outer thread
▪ St. B¼" / St. B½" inner thread
▪ others to agree
Electrical connection
according to DIN43650 or clamp rail
Applicationmeasuring of differential pressure of liquids and gases

                               Look at technical specification

Larger
PXWD series
Product description:
PXWD series is a group of transducers from PXD series wich are equipped in LED digital indicator wich enables direct readout in both pressure or output signal units.

Technical parameters:
Measure
ranges [bar]
▪ 0÷0,050 minimal
▪ 0÷20 maximal
Power supply
[VDC]
▪ 18,5÷36 for 4÷20 mA output signal/current loop
▪ 15÷30 for all output three wires signals
Output signals▪ 4÷20 mA
▪ 0÷20 mA
▪ 20÷0 mA
▪ 0÷10 V
▪ 0÷5 V
Basic error [%MR]▪ ≤ 0,5
▪ ≤ 0,25
Indicator▪ For 4÷20 mA output signal
  - LED, red, 4 digit, 10 mm
  - error 0,1%
  - possibility to set all indications in range from
     9999 to -999
  - possibility to set a reverse indication in
     relation to analog signal ( fe. pressure and
     analog signal down, indication up)
▪ For other output signals
  - LED, red, 3½ digit, 10 mm
Pressure connectors▪ M20x1,5 / M12x1,5 outer thread
▪ G¼" / G½" outer thread
▪ St. B¼" / St. B½" inner thread
▪ others to agree
Electrical connection
according to DIN43650
Applicationmeasuring of differential pressure of liquids and gases

                               Look at technical specification

Larger
MCD series - manometers with digital display
Product description:
MCD series is a group of digital manometers to measure differential pressure constructed using PXD series. Differential digital manometers MCD are equipped in LCD indicator 3,5 digit and own battery power supply. Readout is done during 10 seconds after pressing a "POMIAR" key.

Technical parameters:
Measure
ranges [bar]
▪ 0÷0,050 minimal
▪ 0÷20 maximal
Power supplyalkaline battery 9V
Display kindLCD 8mm, 3.5 digit
Basic error [%MR]▪ ≤ 0,5
▪ ≤ 0,25
Pressure connectors▪ M20x1,5 / M12x1,5 outer thread
▪ G¼" / G½" outer thread
▪ St. B¼" / St. B½" inner thread
▪ others to agree
ApplicationDifferential pressure gases and liquid measurement

                               Look at technical specification
                                                  Look at owner's manual

Larger
PTD series
Product description

PTD series is a group of transducer used in measurement of differential pressure. DC 4mA current supply is necessary, max voltage output signal is obtained from 80 to 1000mVDC. To obtain standard output signal 4-20mA, 0-20mA, 0-10V (0-5V) this series work together with series MPC.


Technical parameters:
Measure
ranges [bar]
▪ 0÷0,050 minimal
▪ 0÷20 maximal
Power supplycurrent to 4 mA
Output signalfrom 0÷80 mV to 0÷1000 mV
Basic error [%MR]▪ ≤ 0,5
▪ ≤ 0,25
Pressure connectors▪ M20x1,5 / M12x1,5 outer thread
▪ G¼" / G½" outer thread
▪ St. B¼" / St. B½" inner thread
▪ others to agree
ApplicationDifferential pressure of gases and liquid measurement

                               Look at technical specification
 

Pressure transducers
Differential pressure transducers

 NPXD series
 PXD series
 PXWD series
 Digital manometers - MCD
 PTD series

Flange separators
Can separators
Tubus separators
Food separators
Level probes
Manometers
Pressure gauges
Equipment

Copyright: Peltron 2005 | created by OrangeDot